I have the following equipment

NI 9234, cDAQ-9188 and an accelerometer PCB 352C03.

My question is,  do I need signal conditioner or charge amplifier to connect accelerometer to NI 9234 or is it possible to connect accelerometer directly to Input Module?

 

Regards

Din

0 Kudos
Message 1 of 3
(2,360 Views)
Solution
Accepted by topic author NAVD

Hello NAVD,

 

do I need signal conditioner or charge amplifier to connect accelerometer to NI 9234 ?

No, You don't need.

is it possible to connect accelerometer directly to Input Module ?

Yes, it is possible. that's the right way.

 

NI 9234 Content all requiement for signal conditionning.

http://www.ni.com/pdf/manuals/374238a_02.pdf

 

Best Regards,

Fred

 